The Transient Hot Wire (THW-L1) is widely used for the accurate measurement of thermal conductivity, thermal diffusivity, and specific heat capacity of liquids, pastes, and powders.
Bitte geben Sie Ihre E-Mail-Adresse ein, um die Broschüre herunterzuladen.
The THW-L1 following ASTM D7896-19, is widely used for the accurate measurement of thermal conductivity, thermal diffusivity and specific heat of liquids and pastes. This versatility is greatly expanded with the addition of Thermtest’s proprietary Temperature Platform (TP) which is appreciated by academic and commercial users alike.
The transient hot wire (THW) has been used for the testing of liquids for more than 30 years, making it one of the most published methods for measuring thermal conductivity of liquids available.
The Transient Hot-Wire (THW-L1) liquid thermal conductivity equipment is an advanced measurement system for testing the thermal conductivity, thermal diffusivity, and specific heat capacity of liquids, pastes and powders in accordance with ASTM D7896-19. The THW-L1 was designed with speed and operational simplicity in mind. With a single measurement of 1 second in duration, small volumes of liquids, pastes, and powders can be accurately and precisely measured for thermal conductivity, thermal diffusivity, and specific heat. The THW-L1 utilizes a non-stationary measurement approach and rapid test times to limit convective effects of samples with a wide range of viscosities.
The THW sensor consists of a thin heating wire 40 mm in length and is completely inserted into the sample to be tested. The sensor wire is heated using a constant current source (q) and the temperature rise is recorded by monitoring the change in electrical resistance of the wire. The slope (m) from the plot of temperature rise vs. logarithm of time is used in the calculation of thermal conductivity (λ). For liquid samples of high thermal conductivity, the lower the slope. For liquid samples of low thermal conductivity, the higher the slope.
Materialien | Flüssigkeiten, Pasten und Pulver |
---|---|
Messfähigkeiten | Flux Sensors (x2) |
Wärmeleitfähigkeit | 0,01 bis 2 W/m•K |
Zusätzliche Eigenschaften | Temperaturleitfähigkeit, und Spezifische Wärme |
Viscosity Range | 0.001 to 10,000,000 cP |
Messzeit | 1 Sekunde | < 5 Sekunden |
Reproduzierbarkeit | ± 1% |
Genauigkeit* | ± 5% |
Temperaturbereich† | -150 to 300°C |
Druck | Up to 35 bar |
Smallest Volume | 20 mL |
Normen | ASTM D7896-19 |
Modelle | THW-L1 | THW-L1S | THW-L1E |
---|---|---|---|
Materialien | Flüssigkeiten, Pasten und Pulver | Flüssigkeiten, Pasten und Pulver | Flüssigkeiten, Pasten und Pulver |
Messfähigkeiten | Schüttgut-Eigenschaften | Schüttgut-Eigenschaften | Schüttgut-Eigenschaften |
Wärmeleitfähigkeit | 0,01 bis 2 W/m•K | 0,01 bis 2 W/m•K | 0,01 bis 2 W/m•K |
Zusätzliche Eigenschaften | Temperaturleitfähigkeit, und Spezifische Wärme | Temperaturleitfähigkeit, und Spezifische Wärme | Temperaturleitfähigkeit, und Spezifische Wärme |
Viscosity Range | 0.001 to 10,000,000 cP | 0.001 to 10,000,000 cP | 0.001 to 10,000,000 cP |
Messzeit | 1 Sekunde | < 5 Sekunden | 1 Sekunde | < 5 Sekunden | 1 Sekunde | < 5 Sekunden |
Reproduzierbarkeit | ± 1% | ± 1% | ± 1% |
Genauigkeit* | ± 2% | ± 2% | ± 2% |
Temperaturbereich | 10 to 200°C | -50 to 200°C | -150 to 300°C |
Druck | Up to 20 bar | Up to 20 bar | Up to 35 bar |
Smallest Volume | 20 mL | 20 mL | 20 mL |
Normen | ASTM D7896-19 | ASTM D7896-19 | ASTM D7896-19 |
Externes Kühlgerät erforderlich | Nein | Ja | Ja |
*All performance data is verified with ASTM Type II Water (IAPWS)
Modelle | THW-L2 | THW-L3 | THW-S |
---|---|---|---|
Materialien | Flüssigkeiten, Pasten und Pulver | Flüssigkeiten, Pasten und Pulver | Isolierung und weiche Materialien |
Messfähigkeiten | Schüttgut-Eigenschaften | Schüttgut-Eigenschaften | Schüttgut-Eigenschaften |
Wärmeleitfähigkeit | 0,01 bis 2 W/m•K | 0,01 bis 1 W/m•K | 0,01 bis 2 W/m•K |
Zusätzliche Eigenschaften | Nein | Nein | Nein |
Viscosity Range | 0.001 to 10,000,000 cP | 0.001 to 10,000,000 cP | K.A |
Messzeit | 1 Sekunde | < 5 Sekunden | 1 Sekunde | 1 to 5 seconds |
Reproduzierbarkeit | ± 2% | ± 2% | ± 2% |
Genauigkeit* | ± 5% | ± 5% | ± 5% |
Temperaturbereich | -50 to 100°C | 10 bis 40°C | 10 bis 40°C |
Druck | Umgebung | Umgebung | K.A |
Smallest Volume | 15 mL | 15 mL | Min. Sample Size: 50mm x 10 mm |
Normen | ASTM D7896-19 | ASTM D7896-19 | K.A |
Externes Kühlgerät erforderlich | Ja | Ja | K.A |
Approximate Time: < 1 min.
Approximate Time: < 1 min.
Approximate Time: 1 min.
Low back pressure was applied to the fluids, to allow measurements over their boiling point.
Wärmeleitfähigkeit deviations from reference values:
Water < 2%
Thermische Diffusivität deviations from reference values:
Water < 2%
Wärmeleitfähigkeit deviations from reference values:
Heptane < 3%
Thermische Diffusivität deviations from reference values:
Heptane < 2%
Calculated Specific Heat Capacity deviations from reference values:
Water < 2%
Heptane < 5%
Wärmeleitfähigkeit deviations from reference values:
Ethylene Glycol < 1%
Thermische Diffusivität deviations from reference values:
Ethylene Glycol < 1%
Calculated Specific Heat Capacity deviations from reference values:
Ethylene Glycol < 2%
Citation:
Lemmon, E.W., Bell, I.H., Huber, M.L., McLinden, M.O. (2018). NIST Standard Reference Database 23: Reference Fluid Thermodynamic and Transport Properties-REFPROP, Version 10.0. National Institute of Standards and Technology, Standard Reference Data Program, Gaithersburg.
Modelle | ||||
Materialien | Flüssigkeiten, Pasten und Pulver | Flüssigkeiten, Pasten und Pulver | Flüssigkeiten, Pasten und Pulver | Flüssigkeiten, Pasten und Pulver |
Module testen | Masse | Masse | Masse | Masse |
Andere Materialien | Feststoffen mit TPS | Nein | Nein | Feststoffen mit TPS, TLS |
Wärmeleitfähigkeit | 0,01 bis 2 W/m•K | 0,01 bis 2 W/m•K | 0,01 bis 2 W/m•K | 0,01 bis 1 W/m•K |
Stichprobengröße | 20 mL | 20 mL | 15 mL | 15 mL |
Zusätzliche Eigenschaften | Temperaturleitfähigkeit, und Spezifische Wärme | Temperaturleitfähigkeit, und Spezifische Wärme | Nein | Nein |
Temperaturbereich | Raumtemperatur 10 bis 200°C | -50/-15/0 bis 200°C -160/-45/-15/0 bis 300 °C | 10 bis 200°C | -50/-15/0 bis 200°C -160/-45/-15/0 bis 300 °C | -50 bis 100 °C | Raumtemperatur 10 bis 40°C |
Druck | Bis zu 35 bar* | Bis zu 35 bar* | Umgebung | Umgebung |
Test Zeiten | 1 Sekunde | 1 Sekunde | 1 Sekunde | < 5 Sekunden | 1 Sekunde |
Daten-Punkte | 400 Punkte/Sekunde | 100 Punkte/Sekunde | 60 Punkte/Sekunde | 60 Punkte/Sekunde |
Genauigkeit der Wärmeleitfähigkeit | 2 % | 2 % | 5 % | 5 % |
Reproduzierbarkeit | 1 % | 1 % | 2 % | 2 % |
Normen | ASTM D7896-19 | ASTM D7896-19 | ASTM D7896-19 | ASTM D7896-19 |
Externes Kühlgerät erforderlich | Nein | Ja* | Nein | Ja* | Ja | Ja |
Füllen Sie einfach das Formular aus und schicken Sie es ab. Eines unserer qualifizierten Teammitglieder wird sich mit Ihnen in Verbindung setzen und Ihnen ein auf Ihre Bedürfnisse zugeschnittenes Angebot unterbreiten.