HFM-25 Débitmètre massique thermique

The HFM-25 is a portable heat flow meter for the measurement of thermal resistance and thermal conductivity of small samples of insulation and construction materials.

Idéal pour small samples of insulation and construction materials
HFM-25 heat flow meter
The Heat Flow Meter (HFM) 25 for the measurement of thermal resistance and thermal conductivity of small samples of insulation and construction materials. Offering the testing power of a full-size heat flow meter, optimized for small insulation samples and budget.

Caractéristiques

HFM-25 features

The economic HFM-25 was designed to meet the growing need for steady-state testing of small samples of insulation for true thermal resistance and thermal conductivity.

Designed with simplicity in mind, a sample is positioned between the two heating – cooling plates. Plate contact is controlled manually, with variable pressure applied with the use of standard weights. Thickness is conveniently reported with included digital gauge. Testing on compressible samples is possible with included shims.

Ambient room temperature is automatically detected, users can choose from 5, 10, 15, and 20 °C delta ranges. One heat flux sensor is integrated into each plate and is used to monitor heat flux (Q/A), generated due to the difference in temperature (ΔT) between the top and bottom plate at regular intervals until steady-state heat flux is observed. The composite heat flux is then used to measure thermal resistance (R) and calculate thermal conductivity (λ) according to Fourier’s Law.

HFM-100 equation

Specification

Conductivité thermique

0,01 à 0,3 W/m•K

Directions

Traversée de l'épaisseur

Précision

3%

Reproductibilité

1%

Température

Delta 5°C, 10°C, 15°C and 20°C at ambient temperature

Type of Sensors

Flux Sensors (x2) with surface thermocouples

Taille minimale de l’échantillon

50 x 50 x 2 mm

Taille maximale de l’échantillon

25.4 mm

Maximum Sample (LxW)

Illimité

Standard

*ASTM C518, C1784, ISO 8301, JIS A1412, and EN 12664

* Applies for materials below 0.1 W/mK

Applications

NIST 1450d high-density fibrous glass board

Location Testing for Homogeneity

The homogeneity of NIST 1450d high-density fibrous glass board was measured with the HFM-25. Designed to test small samples or locations of 25 mm x 25 mm. The open concept sample clamping allows testing of large samples.

The four corners of the board were measured to analyze the homogeneity of sample. The % RSD from location to location was less than 1%, confirming the homogeneity

Aerogel Blanket

Aerogel Blanket

Aerogel blanket sample (50 x 50 x 9.58 mm) was selected to test by the HFM-25. Thermal conductivity of 0.014 W/m•K was measured at 23°C, which is within 3% of the manufacturer’s specification (measured by GHP – ASTM C177).
NIST SRM 1450d

NIST SRM 1450d

A sample of NIST SRM 1450d was measured for thermal conductivity for accuracy. The certified thermal conductivity for the NIST piece at 25°C is 0.0329 W/m•K. The measured thermal conductivity was 0.0330 W/m•K, which is within 1% of the certified value.

Heat Flow Meter (HFM) Models Comparison


HFM-100

HFM-25

Modèles

HFM-100

HFM-50

HFM-25

Matériaux

Insulation, Solids, Construction Materials

Insulation, Solids, Construction Materials

Insulation, Textiles

Type of Sensors

Flux Sensors (x2)

Flux Sensors (x2)

Flux Sensors (x2)

Applications

General Testing

General Testing

Small Insulation

Direction

Through-thickness

Through-thickness

Through-thickness

Plage de conductivité thermique

0.002 to 0.5 W/m•K

0.002 to 0.5 W/m•K

0,01 à 0,3 W/m•K

Chaleur spécifique

Optional

Optional

N/A

High Thermal Conductivity Kit

Up to 2.5 W/m•K

Up to 2.5 W/m•K

N/A

Temps de mesure

30 to 40 minutes

30 to 40 minutes

20 minutes

Reproductibilité

± 0.5 to 1%

± 0.5 to 1%

1%

Précision

± 1 to 2%*

± 1 to 2%*

3%

Plate Temperature Range

-20 to 70 °C

-20 to 70 °C†

De 10 à 40°C

Smallest Sample Size

250 x 250 x 5

150 x 150 x 5

50 x 50 x 2

Taille maximale de l’échantillon

300 x 300 x 100

200 x 200 x 50

300 x 300 x 25

Normes

ASTM C518, C1784, ISO 8301, JIS A1412, EN 12667, and EN 12664

*Performance verified with NIST 1450d / 1450e | Chilled circulator required